The 2 channel Spectral Detector (SD) employs high efficiency Volume Phase Holographic (VPH) transmission grating technology and variable bandpass filters with 1 - 100 nm bandwidth from 400 - 800 nm detection. The hand-picked multi-alkali PMTs provide broad dynamic range for detection up to 800 nm.
The TruSpectral Detection system employs VPH technology, ensuring higher efficiency in fluorescence detection than a reflection type grating based system.
Multi-alkali PMT
High-performance multi-alkali PMT ensures excellent detection efficiency and broad dynamic range with reliable and proven PMT technology.
Multi Channel Lambda Scanning
The new TruSpectral Detection system features two independent spectral detection channels, each configured with a diffraction grating and variable slit for high-resolution wavelength separation and high-speed bandwidth selection. A linear resolution of 2 nm throughout the visible spectrum (400 - 800 nm) enables accurate spectral unmixing of overlapping fluorescence emissions. Two different spectral deconvolution algorithms - normal and blind deconvolution
- quickly and efficiently separate different but overlapping fluorochromes within the specimen.
Variable Barrier Filter (VBF)
With the TruSpectral Detection system, the ideal wavelength spectrum is automatically selected once a fluorochrome combination has been selected using the software. Users can manually adjust the wavelength range according to the fluorescence emission wavelength peaks.