Observation Methods
|
Luminescence imaging, transmitted brightfield, basic transmitted fluorescence
|
|---|
|
LV200 Main Body
|
Light-tight dark box
|
|---|
|
Manual objective lens focusing
|
|
Coaxial XY stage
|
|
Motorized exciting filter wheel with 3 positions for standard 25 mm optical filters
|
|
Motorized emission filter wheel with 6 positions for standard 25 mm optical filters
|
|
Condenser for transmission brightfield coupled to light guide
|
|
C-mount for camera
|
|
Tube lens optimized for luminescence imaging, 0.2 X magnification
|
|
Illumination Unit
|
External halogen lamp housing coupled with light guide
|
|---|
|
Controller
|
IX-UCB controller for filter wheels and illumination
|
|---|
|
Hand Switch
|
Hand switch for filter wheels and illumination
|
|---|
|
Environmental Control
|
Double-layered chamber type incubator for 35 mm dish including controller, stage heater, top cover heater,
|
|---|
|
main body heater, objective heater, flow meter for 5 % CO2, 95 % air
|
|
System Table
|
Approx. 600 (W) x 600 (D) mm x 700 (H) (needed if large-size camera is used)
|
|---|
|
Camera Options
|
Depending on application and required sensitivity, cooled CCD cameras, EM-CCD cameras or deep-cooled slow-scan
|
|---|
|
CCD cameras can be used. (Please ask about camera/software compatibility.)
|
|
Power Consumption
|
Main unit 850 VA
|
|---|
|
Controller 840 VA
|
|
Installation Space
|
Approx. 1500 (W) x 750 (D) mm (depending on configuration)
|
|---|
|
Total Mass Weight
|
Approx. 70 kg (depending on configuration)
|
|---|
|
NA
|
WD (mm)
|
Correction (mm)
|
|---|
|
UPLSAPO 10X
|
0.4
|
3.1
|
0.17
|
|---|
|
UPLSAPO 20X
|
0.75
|
0.6
|
0.17
|
|---|
|
UPLSAPO 40X
|
0.9
|
0.18
|
0.11 - 0.23
|
|---|
|
UPLSAPO 60XO
|
1.35
|
0.15
|
0.17
|
|---|
|
UPLSAPO 100XOI
|
1.4
|
0.13
|
0.17
|
|---|
|
LUCPFLN 20X
|
0.45
|
6.6 - 7.8
|
0 - 2
|
|---|
|
LUCPFLN 40X
|
0.6
|
2.7 - 4
|
0 - 2
|
|---|
|
PLAPON 60XO
|
1.42
|
0.15
|
0.17
|
|---|
Not Available in Your Country