Free replacement of SZ2-STP Arm Used with Microscopes for Equipment Mounting
SZ2-STP Arm Used with Microscope for Equipment Mounting was shipped with the holding parts of the stereo microscope zoom body installed upside down. We would like to therefore offer replacement of the subject product free of charge.
Important Notice Regarding Colony Count & Size Analysis of Incubation Monitoring Systems CM30 and CM20
We have found a problem with our incubation monitoring systems CM30 and CM20 in which the results of the Colony Count & Size analysis differ from the actual measured values. Please find information for affected products and software.
Software Vulnerabilities Notice Regarding Evident Microscope Software
Due to a severe security vulnerability issue in a third-party component used in Evident Application Software, it is strongly advised to install the provided Service Update. This Service Update provides a version of the affected component that closes the vulnerability gap.
Upgrade from Windows 8.1 to Windows 10
Microsoft ended their service support of Windows 8.1 in January 10, 2023. If you are currently using Windows 8.1 with any of our products, we recommend upgrading your PC to Windows 10.
Compatible Versions of Windows 10 with FLUOVIEW FV3000 / FVMPE-RS
Information for the users of FLUOVIEW FV3000 / FVMPE-RS equipped with controller unit with Windows10.
Notice: Measurement Error on cellSens Software V1.18 and Earlier
We recently learned that when cellSens software version V1.18 and earlier is used with the DP74 microscope digital camera, the measurement of a live image may differ from measurements saved on a snapshot of the same image by approximately 20% under certain imaging conditions.
Software Vulnerabilities Notice Regarding Olympus Microscope Software
Due to a serious security vulnerability issue in a third-party component used in Olympus application software, it is strongly advised to install the provided Service Update. This Service Update provides a newer version of the affected component that closes the vulnerability gap.
Vulnerability Notice Regarding Olympus Microscope Software
Due to a severe security vulnerability issue in a third-party component used in Olympus application software, it is strongly advised to install the provided Service Update. This Service Update provides a version of the affected component that closes the vulnerability gap.
针对cellSens软件和独立显微镜相机软件(DP2-SAL)客户的重要更新
Both cellSens software and the DP2-SAL standalone microscope camera software have length measurement functions. Because the software is often shared between multiple users, if someone inadvertently changes the software’s measurement setting, you may not get the correct length measurement results.
关于显微镜数码相机独立连接套件DP2-SAL所使用记录介质的通知
从Windows 7升级到Windows 10
Microsoft will end their service support of Windows 7 in January of 2020. If you are currently using Windows 7 with any of our products, we recommend upgrading your PC to Windows 10.
免费更换偏光显微镜的Bräce-köhler补偿器‘U-CBR1/U-CBR2’
Regarding our product of Bräce-köhler compensator ‘U-CBR1 / U-CBR2’, it was identified that the orientation setting of Gamma axis (Gamma axis : direction of slow optical vibration) was different from the indication of the products when shipping. We would therefore like to replace the applicable products free of charge.
DP25相机商用PC计算机的运行测试结果
DP72相机商用PC计算机的运行测试结果
DP21相机商用PC计算机的运行测试结果
[重要]关于共聚焦激光扫描显微镜等生物显微镜所用奥林巴斯激光设备的重要安全公告
Dear Valued Olympus Customer, Thank you for your continued business with Olympus Corporation. As part of our efforts for continuous improvement and the safe use and operation of our products, we would like to inform you of the following announcement related to your Laser devices for biological microscopes.
[重要]致使用Windows Vista/XP睡眠模式的DP25显微镜数码相机用户